Model Number E4115
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device began taking samples without any button pushed.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the serial number for the device was provided, a review of the device history records will be performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: 12/2099).
 
Event or Problem Description
It was reported that the device began taking samples without any button pushed.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a device history record and manufacturing review was not required as the event was determined to be expected and the investigation did not identify any manufacturing and/or service-related issues.Investigation summary: the encor enspire breast biopsy system was received at the bd-bard global service & repair.The encor enspire breast biopsy system was visually inspected upon receipt and was found to be in poor physical condition.Also, the pin on the top of the monitor arm is sticking when pressing it down and does not secure the riser tube, the small allen screw holding the monitor arm is broken inside of the riser tube, the tray and panels have chipped paint, the right-side panel is broken at the screw hole in the front of the panel and does not secure the panel to the unit, the right-side panel is missing a screw, the rinse pivot is nicked and interfering with the vent plunger, the rinse, vent, and vacuum plunger tips are worn and have saline residue, te rinse, vent, and vacuum bushings are worn and out of round, the silicone tubing is discolored, the cartridge filter is dirty and the manufacturers serial label is not the current revision were found.The device was functionally tested and passed the test as the device operated properly without self-activating.No other anomalies were identified.Therefore, investigation is determined to be unconfirmed for the reported unintended movement issue.The root cause the reported unintended movement issue cannot be determined as the problem could not be reproduced.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 12/2099), g3 h11: h6 (method, result, conclusion) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR ENSPIRE SYSTM
Common Device Name
REUSABLE BIOPSY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14447028
Report Number2020394-2022-00383
Device Sequence Number4310761
Product Code KNW
UDI-Device Identifier00801741086113
UDI-Public(01)00801741086113
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K111100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE4115
Device Catalogue NumberE4115
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer06/20/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1