H10: manufacturing review: a device history record and manufacturing review was not required as the event was determined to be expected and the investigation did not identify any manufacturing and/or service-related issues.Investigation summary: the encor enspire breast biopsy system was received at the bd-bard global service & repair.The encor enspire breast biopsy system was visually inspected upon receipt and was found to be in poor physical condition.Also, the pin on the top of the monitor arm is sticking when pressing it down and does not secure the riser tube, the small allen screw holding the monitor arm is broken inside of the riser tube, the tray and panels have chipped paint, the right-side panel is broken at the screw hole in the front of the panel and does not secure the panel to the unit, the right-side panel is missing a screw, the rinse pivot is nicked and interfering with the vent plunger, the rinse, vent, and vacuum plunger tips are worn and have saline residue, te rinse, vent, and vacuum bushings are worn and out of round, the silicone tubing is discolored, the cartridge filter is dirty and the manufacturers serial label is not the current revision were found.The device was functionally tested and passed the test as the device operated properly without self-activating.No other anomalies were identified.Therefore, investigation is determined to be unconfirmed for the reported unintended movement issue.The root cause the reported unintended movement issue cannot be determined as the problem could not be reproduced.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 12/2099), g3 h11: h6 (method, result, conclusion) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|