• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 8MM LENGTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 8MM LENGTH Back to Search Results
Model Number CM318
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the inner vial was deteriorated, in bad condition and implant was bit placed.Procedure was completed.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).One implant was returned for investigation.Visual evaluation of the as returned product identified no damage or signs of malfunction that would contribute to the event.Measurements match drawing.(b)(6) (due: (b)(6), 2022) functional testing to recreate the reported event could not be performed due to the nature of the device & event.A pre-existing condition noted on the per was low bone density (type iv).The reported device was intended for tooth # 12 (fdi) and was not used.The customer did not provide any pictures or x-rays.Review of appropriate documentation: documents reviewed: instructions for use for implants, abutments and restorative components for 3.1mmd eztetic¿ implant ifu 9228 rev.1 ¿ 11/19.Information identified: 'contraindications' 'warnings' 'precautions' 'adverse effects'.Dhr review: dhr review was completed for the subject lot number (1240108).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (1240108) for similar events and no other complaint was identified.Review completed utilizing keywords: 'damaged or un-sealed sterile barrier' post market trend review: august post market trending was reviewed and there were no actionable events or corrective actions for the reported event (damaged or un-sealed sterile barrier) or product (cm318).No actionable items have been triggered that will affect complaint handling on our end for this month.Based on the available information, device malfunction did not occur and the reported event was non-verifiable.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 8MM LENGTH
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14447286
Report Number0002023141-2022-01230
Device Sequence Number4576789
Product Code DZE
UDI-Device Identifier00889024343627
UDI-Public(01)00889024343627(17)251008(10)1240108
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K142082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2025
Device Model NumberCM318
Device Catalogue NumberCM318
Device Lot Number1240108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer10/24/2022
Initial Report FDA Received Date05/19/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexMale
-
-