• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 32 + 4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 32 + 4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 71303204
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Joint Dislocation (2374); Metal Related Pathology (4530)
Date of Event 03/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).Mastel, m., boisvert, a., moore, r., sutherland, f., & powell, j.(2022).Metallosis following hip arthroplasty: two case reports.Journal of medical case reports, 16(1), 1-9.Doi: https://doi.Org/10.1186/s13256-022-03336-4.
 
Event or Problem Description
"it was reported that on literature review ""metallosis following hip arthroplasty: two case reports"", one patient had a primary right hip tha surgery with a r3 acetabular shell, r3 poly liner, synergy stem and cocr femoral head.8 years after the primary surgery, the patient suffered from a dislocation and one year later, the patient sustained a second dislocation.The patient exhibited signs of metallosis and underwent a revision surgery to treat the adverse event.At 6 weeks follow-up, the patient was recovering well.
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that no conclusions can be made from this publication as its limited to the information provided and further investigation is not possible.It was reported that at the 6 weeks follow-up, the patient was recovering well.No further clinical assessment is warranted at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems reveal in the postoperative section that patients should be warned against unassisted activity, particularly activities requiring excessive motion of the hip, as they may result in dislocation.Also, although rare, metal sensitivity reactions have been reported in patients following joint replacement, this has been identified in the adverse events in primary and revision surgery section.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy, abnormal loading of limb, damaged product, implant corrosion or wear.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HEAD 32 + 4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14450385
Report Number1020279-2022-02439
Device Sequence Number6569958
Product Code LPH
UDI-Device Identifier03596010194190
UDI-Public03596010194190
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K963509
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Literature,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71303204
Device Catalogue Number71303204
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer06/10/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date06/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKN SYNERGY HIP STEM CEM
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-