COVIDIEN NANOCROSS ELITE 0.14 OTW PTA DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
|
Back to Search Results |
|
| Model Number |
AB14W050120150 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Physician used a nanocross elite pta balloon during treatment of a fibrous cto (chronic total occlusion-100%) in the patient's proximal, mid, and distal superficial femoral artery (sfa).Severe vessel tortuosity and slight calcification are reported.Isr is reported.A 6mm spider fx embolic protection device was used.A non-medtronic inflation device was used for balloon inflation.50/50 contrast to saline was used for inflation.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.No issues were noted when removing the device from the hoop/tray.The device was prepped as per ifu without issue.The device was passed through a previously deployed stent.No resistance was encountered when advancing the device.A longitudinal balloon burst is reported at 8atm during balloon inflation.There were no prior inflations applied to balloon.The balloon did not fragment.No vessel injury noted, balloon was inflated inside stent, stent was not damaged.The physician pulled a second balloon and completed the case.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Image analysis:the customer returned one image.The image shows the shelf carton label.The label indicates that the device is a nanocross elite 5 x 120mm (ab14w050120150), lot #: b034318.Product analysis: the device was returned with a longitudinal tear on the balloon.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|