• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PNEUMOCLEAR PLUS CO2 CONDITIONING INSUFFLATOR (AMERICAS); INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PNEUMOCLEAR PLUS CO2 CONDITIONING INSUFFLATOR (AMERICAS); INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Model Number FM300
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code Cardiac Arrest (1762)
Date of Event 04/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.Filling on behalf of world of medicine (wom).
 
Event or Problem Description
It was reported that the patient coded during procedure and was airlifted to another hospital.
 
Event or Problem Description
It was reported that the patient coded during procedure and was airlifted to another hospital.
 
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.Alleged failure: unknown issue with insufflator.Probable root cause: because device was not returned to oem - wom, probable root cause cannot be determined.A veress needle is used to create pneumoperitoneum for laparoscopic surgery.When the needle is placed in a peritoneal cavity, co2 passed from the insufflator through the tube set and the veress needle to fill the abdomen with gas and create a pneumoperitoneum.In the present case an issue with the insufflator was not observed.After further discussion with our distributor, stryker, who was in contact with the complainant, opined that the physician likely set the veress needle incorrectly and the patient then coded.The currently assumed most possible root cause might be incorrect positioning of the veress needle leading to air embolism.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUMOCLEAR PLUS CO2 CONDITIONING INSUFFLATOR (AMERICAS)
Common Device Name
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14453107
Report Number0002936485-2022-00255
Device Sequence Number20421841
Product Code HIF
UDI-Device Identifier04056702001308
UDI-Public04056702001308
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K170784
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFM300
Device Catalogue Number0620050000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-