| Model Number |
9735542 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The tubing was returned to the manufacturer for evaluation.Analysis found that manipulation of the tubing at the female luer connection point resulted in a leak.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Concomitant medical products: other relevant device(s) are: product id: 9735560, serial/lot #: (b)(4), ubd: 15-feb-2024, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a thermal therapy system being used during a soft neuro tissue ablation procedure.It was reported that the tubing of the thermal therapy system leaked.There was a reported delay to the procedure of five minutes due to this issue.There was no reported impact on patient outcome.
|
| |
|