Model Number 60-02-60
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
Livanova received report that during setup it was confirmed that the evo clamp did not work when the rotation of cp5 stopped.Reportedly, the evo did not close and the link worked when using a roller pump instead of a centrifugal pump cp5.There was no patient involvement.
 
Additional Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova field service representative tested the device and could not reproduce the reported issue.It was reported that the evo was configured correctly at the time of the event.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
H.10: as reported in the event description, the evo was also tested with a roller pump and the stop link function worked properly.Therefore the issue has been addressed to a failure of cp5 panel which was inspected at livanova japan.The issue couldn't be reproduced, however the processor board of the cp5 control panel was replaced.Complaints database review was performed and no further issue has been submitted for this unit since its manufacturing date in 2018.Based on above facts, it cannot be ruled out that the root cause of the reported event (communication issue) can be traced back to an intermittent electrical failure of the processor board.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Common Device Name
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key14453217
Report Number9611109-2022-00253
Device Sequence Number6569438
Product Code DWA
UDI-Device Identifier04033817901006
UDI-Public(01)04033817901006(11)180111
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112225
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number60-02-60
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer06/11/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2018
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-