• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT351418
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 04/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On an unspecified date the patient presented with abdominal/groin pain.Computed tomography angiography (cta) was performed and revealed migration of a previously implanted gore® excluder® aaa endoprosthesis (rlt351418), and a proximal type 1 endoleak.As reported, the previously implanted rlt rlt351418 had migrated 9.5 centimeters and the identified cause was ongoing disease progression with aortic neck dilation.The aneurysm was observed to have grown approximately 8 centimeters in diameter.On (b)(6) 2022 a reintervention was performed.The physician elected to use a gore® tag® conformable thoracic stent graft with active control system (ctag ac, catalog#: tgmr404010) to contain a rupture of the aneurysm.As reported, the ctag ag was used in an unspecified off-label manner.Reportedly, the type i endoleak was contained.The patient was observed to be doing well.The physician does not require any follow-up on this case.No images are available for consideration.
 
Additional Manufacturer Narrative
While there was no device return an evaluation was performed with the available provided information.According to the engineering evaluation: 1) without a physical sample or images to evaluate, the physician¿s observation that there was a type i endoleak, 9.5 centimeter migration, and a rupture of the aneurysm, could not be confirmed.2)the likely cause for the reported type i endoleak, 9.5 cm migration, and rupture of the aneurysm could not be determined with the available information.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.G3/g4 date received by manufacturer was updated to 30-aug-2022, as the additional engineering evaluation results were received on this date.
 
Additional Manufacturer Narrative
This event is both a product malfunction and a serious injury so both have been indicated in section b1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
breanna cox
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14453367
Report Number3007284313-2022-01924
Device Sequence Number11899280
Product Code MIH
UDI-Device Identifier00733132618491
UDI-Public00733132618491
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/03/2019
Device Model NumberRLT351418
Device Catalogue NumberRLT351418
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer08/30/2022
08/30/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date08/30/2022
09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/04/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age89 YR
Patient SexMale
-
-