• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED (PS) NARROW RIGHT SIZE 6; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED (PS) NARROW RIGHT SIZE 6; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Loss of or Failure to Bond (1068); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Synovitis (2094); Swelling/ Edema (4577)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient was revised approximately nine weeks post implantation due to due to pain, swelling, and possible loosening of the tibial component.During revision, no gross loosening was found in the tibial or femoral components.Mild synovitis was noted.The patella was left intact, all other components were revised without complication.Attempts to obtain additional information have been made; however, no more is available.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2022-00134.Medical product tibia cemented 5 degree stemmed right size d, item# 42532006702, lot# 64686661.Articular surface fixed bearing (ps) right 11 mm height, item# 42521400511, lot# 63750891.All poly patella cemented 29 mm diameter, item# 42540000029, lot# 64740327.Biomet bc r 1x40 us, item# 110035368, lot# f2701z129ba.Biomet bc r 1x40 us, item# 110035368.Lot# f2701z129ba.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Primary operative notes do not suggest intraoperative complications.Revision operatives notes state that revision was performed due to pain and questionable tibial loosening.Mild synovitis was noted.Neither tibial nor femoral components were grossly loose.Patella left intact, all other components revised without complication.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.H6: component code: proposed component (annex g) code is mechanical g4- femur.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED (PS) NARROW RIGHT SIZE 6
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14454045
Report Number3007963827-2022-00133
Device Sequence Number15037984
Product Code JWH
UDI-Device Identifier00889024228740
UDI-Public(01)00889024228740(17)290831(10)64472820
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42500006002
Device Lot Number64472820
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/06/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight84 KG
-
-