Other relevant device(s) are: product id: e tcf3636c49e, serial/lot #: (b)(4), ubd: 20-mar-2024, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Product analysis conclusion: the reported loss of seal and proximal endoleak, coverage of the lower pole of the renal artery and interaction between the reliant balloon and left renal artery snorkel balloon leading to snorkel balloon rupture could not be assessed on the pre-implant images provided; therefore, the cause of the events could not be determined.Procedural angiograms showing deployment of the proximal stents and subsequent ballooning of devices were not provided for analysis of the reported events.Endoleak interrogation via selective procedural angiograms (injecting from different levels within the stent graft) to help determine the source and rule out other potential endoleak sources was not available.Post-implant ct¿s showing the stent graft in vivo configuration were also not received.It is most likely that the patient anatomy (short, wide, angled , thrombus filled and conical infrarenal aortic neck) was a factor in the occurrence of the reported proximal type ia endoleak.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|