Model Number APB-1-2-HX-ES
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding an axium coil that detached prematurely.The patient was undergoing a coil embolization procedure to treat a ruptured saccular left internal carotid artery (ica) c4 segment aneurysm.The aneurysm max diameter was 3mm and the neck diameter was 3.5mm.Blood flow and vessel tortuosity were normal.It was reported that the axium coil and all accessory devices were prepared as indicated in the instructions for use (ifu).The coil was delivered with the microcatheter.After filling about 1cm, the physician wanted to make an adjustment.When the pusher was pulled back, there was no resistance and the coil did not follow the movement.Imaging confirmed that the pusher was withdrawn about 10cm and the detachment point of the coil did not change indicating that the proximal end of the coil was detached from the pusher in the microcatheter.Therefore, the pusher was used to push the coil into the aneurysm and the coil was implanted in the intended location.No additional medical or surgical intervention was needed.It was noted that there was no friction or other difficulty during delivery of the coil.The pusher was not bent or broken.Only one attempt was made to reposition the coil.The physician did not attempt to detach the coil and did not rotate the delivery pusher.Continuous flush was administered.There was no harm or injury to the patient.Ancillary devices: cook 6f long sheath, headway 17 microcatheter, synchro14 guidewire.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL HLX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key14454318
Report Number2029214-2022-00852
Device Sequence Number6418921
Product Code KRD
UDI-Device Identifier00847536032095
UDI-Public00847536032095
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/12/2024
Device Model NumberAPB-1-2-HX-ES
Device Catalogue NumberAPB-1-2-HX-ES
Device Lot NumberB189737
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date06/14/2022
Date Device Manufactured04/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age37 YR
Patient SexMale
Patient Weight70 KG
-
-