Model Number 25-302
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Stroke/CVA (1770)
Date of Event 05/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
This medwatch is not to report a device malfunction, but to report an adverse patient effect post procedure.An angiodynamics territory manager (tm) reported via a product evaluation survey, a patient adverse event occurred after treatment with an alphavac multipurpose mechanical aspiration f18-85 cannula.The patient was treated for a left atrial thrombus, removing approximately 95% of the thromboemboli, and no issues were noted during the procedure.However, approximately 2-3 hours after the procedure was completed, the patient suffered a stroke.At this time the patient is currently stable and improving.Follow up with the tm determined that the physician confirmed this was not, in anyway, a result of product malfunction but a complication of the procedure.
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The product was not returned for evaluation since there was no report of alphavac malfunction during the procedure; just patient serious adverse event (stroke).The customer's reported complaint description cannot be confirmed due to the nature of this patient serious adverse event; there were no reports of alphavac mma gen 3 device malfunction during the procedure.No sample was returned for evaluation.It was concluded that the alphavac device was unrelated to the patient's stroke (distal embolization of thrombus); i.E.This was just an anticipated procedural complication.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: directions for use is provided with this device and contains the following statements: adverse events this device, as do all embolectomy systems and devices, has possible side effects, which include, but are not limited to infections, blood loss, thrombus formation, embolic events, vessel, ventricular or valvular damage and complications of percutaneous or surgical insertion techniques.These may occur if the directions for use are not followed.Possible complications include those normally associated with large bore surgical and/or percutaneous vessel catheterization/cannulation, anticoagulation, and application of intravascular introducer systems which include but are not limited to: · death · distal embolization of thrombus · vascular thrombosis a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALPHAVAC
Common Device Name
ALPHAVAC MULTIPURPOSE MECHANICAL ASPIRATION 25F/C20
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key14454360
Report Number1319211-2022-00022
Device Sequence Number17727554
Product Code QEZ
UDI-Device IdentifierH787253020
UDI-PublicH787253020
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K211081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/27/2023
Device Model Number25-302
Device Catalogue Number725302
Device Lot Number420017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight76 KG
-
-