C.R. BARD, INC. (COVINGTON) -1018233 SURESTEP INTERMITTENT CATHETER TRAY Back to Search Results
Model Number INTS16
Medical Device Problem Code Material Too Rigid or Stiff (1544)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the customer recently converted from ureteral catheter tray to intermittent catheter tray and said that the intermittent catheter tray was stiffer and had resulted in patient injury by penetrating the urethra.No medical intervention was reported.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the customer recently converted from ureteral catheter tray to intermittent catheter tray and said that the intermittent catheter tray was stiffer and had resulted in patient injury by penetrating the urethra.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "operator error".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Bard® ez-lok¿ sampling port (indicated by the blue stem in the port) accepts luer lock (fig.1a) or slip tip collection devices (fig.1b).Bd vacutainer® urine products can be used for collection, storage, and transport of urine specimens from a urinary catheter.1.Swab surface of bard® ez-lok¿ sampling port with antiseptic (fig.2) e.G., site-scrub¿ ipa device which is an effective disinfecting agent for luer hubs.2.Using aseptic technique, position the collection device in the center of the sampling port.Press the device firmly and twist gently to access the sampling port.(fig.3) 3.If using bd vacutainer® luer-lok¿ access device (fig.4), collect urine into the bd vacutainer® tubes per hospital protocol.If using syringe, slowly aspirate urine sample into syringe and transfer to collection cup or follow hospital protocol.4.After sample is obtained, discard collection device according to hospital protocol.5.Follow established hospital protocol for specimen labeling and transport to lab".H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESTEP INTERMITTENT CATHETER TRAY
Common Device Name
INTERMITTENT CATHETER TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14454620
Report Number1018233-2022-03933
Device Sequence Number20418048
Product Code KOD
UDI-Device Identifier00801741183225
UDI-Public(01)00801741183225
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberINTS16
Device Catalogue NumberINTS16
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer07/18/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-