| Model Number |
IPN000096 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cath kit was defective, and the balloon did not inflate during use on a patient.As a result, a new catheter was used successfully.There was no report of patient complications, serious injury, or death.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).There is an investigation ongoing into this reported complaint.A supplement follow-up mdr report will be submitted with the investigation results upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The product was not returned for investigation.The reported complaint of balloon would not inflate in use is not able to be confirmed.If the product is returned at a later date, a full investigation of the sample will be completed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the cath kit was defective, and the balloon did not inflate during use on a patient.As a result, a new catheter was used successfully.There was no report of patient complications, serious injury, or death.
|
| |
|
Search Alerts/Recalls
|