• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL HD EPSCP,1.9,30,60,CW_STORZ; RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL HD EPSCP,1.9,30,60,CW_STORZ; RIGID ENDOSCOPE Back to Search Results
Model Number 242048
Medical Device Problem Codes Break (1069); Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by sales rep that a hd arthroscope/sinuscope, ep, 1.9mm, 30 deg, 60mm (storz style) device was cracked.During in-house engineering evaluation, it was determined that the outer tube damaged, bent/dented, distal tip was damaged, distal tip fiber damaged, the compact objective was loose, the lenses were broken in the optical system and less than 50% of lenses were defective.There was no procedure nor patient involvement reported.No additional information was provided.
 
Additional Manufacturer Narrative
Additional narrative: udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: outer tube damaged -bent/dented, outer tube damaged, distal tip damaged, distal tip fiber damaged, optical system, optical components compact objective loose, broken lenses in optical system, less than 50% of lenses defective.Per service reports, this complaint can be confirmed.The defective parts needs to be replaced to resolve the issues.As part of mitek's quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD EPSCP,1.9,30,60,CW_STORZ
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle CH-24-00
SZ   CH-2400
3035526892
MDR Report Key14454822
Report Number1221934-2022-01464
Device Sequence Number6354738
Product Code HRX
UDI-Device Identifier10886705028108
UDI-Public10886705028108
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K080560
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number242048
Device Catalogue Number242048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2022
Initial Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-