| Model Number |
242048 |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by sales rep that a hd arthroscope/sinuscope, ep, 1.9mm, 30 deg, 60mm (storz style) device was cracked.During in-house engineering evaluation, it was determined that the outer tube damaged, bent/dented, distal tip was damaged, distal tip fiber damaged, the compact objective was loose, the lenses were broken in the optical system and less than 50% of lenses were defective.There was no procedure nor patient involvement reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: outer tube damaged -bent/dented, outer tube damaged, distal tip damaged, distal tip fiber damaged, optical system, optical components compact objective loose, broken lenses in optical system, less than 50% of lenses defective.Per service reports, this complaint can be confirmed.The defective parts needs to be replaced to resolve the issues.As part of mitek's quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Search Alerts/Recalls
|
|
|