| Model Number |
STRIP, CVS KETONE 50CT |
| Medical Device Problem Code |
Material Discolored (1170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Ketone test strips were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on 05-may-2022 to ensure the replacement product resolved the initial concern - able to establish contact with customer who stated replacement product resolved initial concern.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for the ketone test strips.Customer stated that the package was not opened or damaged when received.Customer stated the test strip vial was sealed and when she opened it, the color on the test strip pad was blue in color instead of beige.Customer did not perform any test using the test strips.The customer feels well and did not report any symptoms and medical attention related to the use of the product was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 26-july-2022: h6: updated fda's type of investigation, investigation findings, and investigation conclusions.H10: ketone test strips were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strips tested within specifications.Most likely underlying root cause: mlc-050: manufacturing or packaging process caused malfunction of user's test strip.
|
| |
|
Search Alerts/Recalls
|