• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST MESH; SURGICAL MESH Back to Search Results
Catalog Number 5954113
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Abdominal Pain (1685); Hernia (2240)
Date of Event 04/01/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported, about 1 year post-implant of the bard/davol ventralight st mesh, patient was diagnosed with hernia recurrence.As reported, no action has been taken and it was not-resolved.The clinician has assessed the patient¿s postoperative course as being ¿possibly¿ related to the study device and the procedure.However, based on the information provided, no conclusion can be made.Hernia recurrence is a known inherent risk of hernia repair surgery.The instructions-for-use supplied with the device lists hernia recurrence as a possible complication.Review of manufacturing records confirms product was manufactured to specification.This mdr represents the ae of hernia recurrence.An additional mdr was submitted to represent an additional ae of seroma.Not returned - remains implanted.
 
Event or Problem Description
Reported per clinical trial (b)(6): on (b)(6) 2013 - subject patient underwent a primary laparoscopic hernia repair and was implanted with a bard/davol ventralight st mesh.The hernia defect was 23 x 13 cm extending from old ostomy site to midline.The mesh was fixated using mechanical fixation (sorbafix) and sutures.Fascia closure was achieved; full skin closure was achieved using adhesive skin closure.The patient was discharged from the hospital on (b)(6) 2013.On (b)(6) 2014 - subject patient presented with abdominal pain and right upper quadrant bulge after coughing fit.The surgeon requested a ct to rule out recurrence; patient did not undergo ct or returned to clinic.On (b)(6) 2014 - patient was diagnosed with incisional hernia recurrence during physical exam.Per the study clinician, the reported adverse event has been assessed as moderate in severity, possibly related to the study device and to the index procedure.The outcome for this ae is reported as not recovered/ not resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14454871
Report Number1213643-2022-00248
Device Sequence Number6418946
Product Code FTL
UDI-Device Identifier00801741031595
UDI-Public(01)00801741031595
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K101851
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source Other,Study,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2014
Device Catalogue Number5954113
Device Lot NumberHUWJ1592
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Initial Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/31/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexMale
Patient Weight119 KG
-
-