• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 3549311
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: defective valve manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) female who underwent an unspecified breast surgery with a mentor cpx 4 breast tissue expander with suture tabs, 250cc breast implant experienced left sided ¿defective valve¿ intraoperatively.The report indicated that moisture was detected on the edge of the expander.No adverse event reported.
 
Additional Manufacturer Narrative
Device evaluation completed on may 23 , 2022: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample determined that a tear was found on the union between the shell and anterior patch aspect of the tall height te wsut tabs 250cc tissue expander.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the tissue expander shell.Based on the information currently available, microscopic examination of the returned product indicates that the tissue expander could have been damaged during to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
646591-7981
MDR Report Key14455123
Report Number1645337-2022-05867
Device Sequence Number17747104
Product Code LCJ
UDI-Device Identifier00081317006770
UDI-Public00081317006770
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
K130813
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2025
Device Model Number3549311
Device Catalogue Number3549311
Device Lot Number9524144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/29/2022
Supplement Date Received by Manufacturer05/23/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
-
-