| Model Number |
3549311 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: defective valve manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) female who underwent an unspecified breast surgery with a mentor cpx 4 breast tissue expander with suture tabs, 250cc breast implant experienced left sided ¿defective valve¿ intraoperatively.The report indicated that moisture was detected on the edge of the expander.No adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on may 23 , 2022: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample determined that a tear was found on the union between the shell and anterior patch aspect of the tall height te wsut tabs 250cc tissue expander.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the tissue expander shell.Based on the information currently available, microscopic examination of the returned product indicates that the tissue expander could have been damaged during to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|