• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HZ APPLIER MICRO 8" CVD; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HZ APPLIER MICRO 8" CVD; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN008965
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device investigation is pending.
 
Event or Problem Description
When closing vessel with microclip vessel is cut.
 
Event or Problem Description
When closing vessel with microclip vessel is cut.
 
Additional Manufacturer Narrative
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet, inc.Kenosha wi facility as part of a (b)(4)-piece lot in may of 2018.It was found that this order was made from the correct materials and components and all approved processes were followed so it can then be stated that the alleged non-conformance inciting this complaint was not due to an error in tecomet - kenosha's manufacturing process.Evaluation of the returned instrument shows that the jaws are aligned with each other in the closed position and no non-conforming features were found.Functional testing shows that this instrument is able to pick-up, retain, close and release multiple clips without any issues.We are unable to replicate the alleged issue therefore we are unable to validate this complaint.All (b)(4) instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HZ APPLIER MICRO 8" CVD
Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key14455162
Report Number3011137372-2022-00080
Device Sequence Number6341030
Product Code GDO
UDI-Device Identifier24026704696639
UDI-Public24026704696639
Combination Product (Y/N)N
Initial Reporter CountryFI
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN008965
Device Catalogue Number537081
Device Lot Number06J1757100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer06/17/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/16/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-