| Model Number |
015914 |
| Medical Device Problem Code |
Expiration Date Error (2528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the user enquired regarding the expiry date of the female catheter product since it had different dates on packaging (05/2016 and 11/2016) and jde system (08/31/2026).Lot # ngfv2870 in jde it showed as 30nov2026 and lot #ngfv2874, in jde showed as 31aug2026.Per additional information received on 04may2022, three different lots were obtained (lot# ngfv2870, lot #ngfv2874, lot #ngez1598).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed because the device meets specifications.The device met relevant specifications/ the reported failure is within specification as the reported failure could not be reproduced.The product was not used for patient treatment or diagnosis.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.Visual evaluation of the returned sample noted a box of 50 unopened intermittent catheters.Visual inspection noted the expiration date on the box (06/2016) does not match the product expiration dates (10/2016).The 2016 dates reported are not expiration dates, but rev.Dates of the pk components.The device history record review was not required as the reported event was unconfirmed.The labeling review was not required as the reported event was unconfirmed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that the user enquired regarding the expiry date of the female catheter product since it had different dates on packaging (05/2016 and 11/2016) and jde system (08/31/2026).Lot # ngfv2870 in jde it showed as 30nov2026 and lot #ngfv2874, in jde showed as 31aug2026.Per additional information received on 04may2022, three different lots were obtained (lot# ngfv2870, lot #ngfv2874 , lot #ngez1598).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.One box of 50 samples exhibited the reported failure.The device had not met specifications.The reported failure is considered out of specification as the reported failure was reproduced.The product was not used for patient treatment or diagnosis.A potential root cause for this failure could be error on selection of character.Visual evaluation of the returned sample noted a box of 50 unopened intermittent catheters.Visual inspection noted the expiration date on the box (06/2016) does not match the product expiration dates (10/2016).This is out of specification per inspection procedure which states, printing must be clear and distinct.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use: 1.Starting at the top of the package, peel down to expose the catheter.2.Holding the labia apart with one hand , take the catheter and insert the tip of the catheter slowly into the urethra.3.If sitting on a toilet seat, once the tip of the catheter is in the bladder, allow the urine to flow directly into the toilet bowl.If on a chair or in bed, use a receptacle for urine collection." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the user enquired regarding the expiry date of the female catheter product since it had different dates on packaging (05/2016 and 11/2016) and jde system (08/31/2026).Lot # ngfv2870 in jde it showed as 30nov2026 and lot #ngfv2874, in jde showed as 31aug2026.Per additional information received on 04may2022, three different lots were obtained (lot# ngfv2870, lot #ngfv2874 , lot #ngez1598).
|
| |
|
Search Alerts/Recalls
|