• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. CREO STABILIZATION SYSTEM; CREO AMP THREADED POLYAXIAL TULIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. CREO STABILIZATION SYSTEM; CREO AMP THREADED POLYAXIAL TULIP Back to Search Results
Model Number 1119.0110
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a revision surgery due to (2) screw heads failing post operatively.
 
Additional Manufacturer Narrative
The device was not available for evaluation as it was discarded by the hospital.The post operative imaging provided from the original surgery shows both the left and right s2 screws heads had dissociated from their screw shanks.The imaging provided of the explanted s2 amp screw heads showed that they were subjected to excess force consistent with head pull-off.One implant was missing its internals and the second implant had internals that were significantly exposed with evidence of lateral deformation.Pseudoarthrosis at l5-s1 was noted during the revision surgery.It is possible the screw head dissociation is a result of excess bending forces on the subject s2 implants resulting from inadequate fixation, both from pseudoarthrosis and the lack of s1 fixation at the time of original implantation.However, an exact cause to the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CREO STABILIZATION SYSTEM
Common Device Name
CREO AMP THREADED POLYAXIAL TULIP
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key14455306
Report Number3004142400-2022-00074
Device Sequence Number5050603
Product Code MNH
UDI-Device Identifier00889095059656
UDI-Public00889095059656
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K124058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1119.0110
Device Lot NumberBAX010DG/BAX153NG
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Initial Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-