MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71335550
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 01/11/2011
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event or Problem Description
It was reported that, during primary left thr, the patient experienced an unspecified intraoperative fracture.The patient needed a subsequent fixation.The location of the fracture was not reported.The outcome of the plaintiff after the primary surgery is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the limited information provided, we are unable to rule out a procedural variance as a contributing factor to the reported intraoperative fracture.Without the supporting primary and revision operative reports and/or imaging, the root cause cannot be confirmed.It cannot be concluded that the reported event was associated with a mal-performance of the implant or implant failure.The patient impact beyond that which has already been reported cannot be determined.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for hip systems revealed that care should be taken intraoperatively to avoid fracture of the component and/or bone have been identified as warnings and precautions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or potential causes that could contribute to the reported event include excessive forces applied to implant and surgical technique.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 50MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14455377
Report Number1020279-2022-02461
Device Sequence Number20418056
Product Code MBL
UDI-Device Identifier03596010598233
UDI-Public03596010598233
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/08/2021
Device Model Number71335550
Device Catalogue Number71335550
Device Lot Number11GM05073
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/11/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-