• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG AS FEMUR EXTENS.STEM 6° D12X77 CEMENTED; KNEE ENDOPROSTHESES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS FEMUR EXTENS.STEM 6° D12X77 CEMENTED; KNEE ENDOPROSTHESES Back to Search Results
Model Number NR291Z
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.Batch history review: the device history records (dhr) were reviewed for the reported lot number; the device was found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.No similar complaints were identified against the same lot number with a similar reported event.Conclusion: a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.There is no indication of a material defect, manufacturing failure, or design error on the basis of the device history record.For further investigation, examination of the complaint device is required.
 
Event or Problem Description
It was reported to aesculap ag that a as femur extens.Stem 6° d12x77 cemented (part # nr291z) was implanted during a left knee revision surgery performed on (b)(6) 2018.According to the complainant, post-operative bone deterioration of femur and tibia occurred.The complaint device has not been returned to the manufacturer for evaluation.An additional medical intervention was required.A revision in the future is likely, according to the patient.The malfunction is filed under aag reference (b)(4).Associated medwatch reports: 400554585 - nr191z (9610612-2022-00142).400554587 - nr565z (9610612-2022-00144).400554588 - nr565z (9610612-2022-00145).400554589 - nr611m (9610612-2022-00146).400554590 - nr071z (9610612-2022-00147).400554591 - nr005z (9610612-2022-00148).400554592 - nr400z (9610612-2022-00149).
 
Additional Manufacturer Narrative
Additional information: b5 - additional information b7 - patient history d6b - explant date h6 - codes updated investigation results: the complained product was not available for investigation.Results were based upon batch history review and event description.Batch history review: the device quality and manufacturing history records have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number.According to manufacturer's reporting evaluation in accordance with 21 cfr part 803, section 803.3, this event is considered reportable for the following reason - adverse event (not later than 30 days).The assessment for the reportability of this adverse event was based on the patient harm, revision.Conclusion/preventive measures: on the basis of the current information and without the product for investigation, a clear conclusion cannot be drawn.There is no indication for a material defect or manufacturing failure on the basis of the device history records.Despite the additional information provided by the patient (submitted previously), a root cause for the problems mentioned cannot be clearly defined.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigation results, a capa is not required.
 
Event or Problem Description
Update: revision surgery was performed on (b)(6)2023; implant had come apart at the post and plate.It was noted that glue had not been the correct type, stated by surgeon.Treatment for 5 years prior to this included bracing, radio frequency ablation procedures, and opiods.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS FEMUR EXTENS.STEM 6° D12X77 CEMENTED
Common Device Name
KNEE ENDOPROSTHESES
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key14456867
Report Number9610612-2022-00143
Device Sequence Number20014217
Product Code KRO
UDI-Device Identifier04046963176464
UDI-Public4046963176464
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K101815
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2024
Device Model NumberNR291Z
Device Catalogue NumberNR291Z
Device Lot Number52046962
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2022
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date04/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/18/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age70 YR
Patient SexFemale
-
-