• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM TI CANN TIBIAL NAIL-EX W/PROX BEND 315MM-STERILE; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM TI CANN TIBIAL NAIL-EX W/PROX BEND 315MM-STERILE; NAIL, FIXATION, BONE Back to Search Results
Model Number 04.034.443S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 04/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional narrative: 510k: this report is for an unknown tibial nail/unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a tibial nail was removed due to infection and non-union.An external fixator will be placed because of a possible infection.The patient did not experience a post-op device malfunction.This nail was originally implanted in (b)(6) 2021.During the reaming process, during the revision, they attempted to bend the tip of a 2.5 reaming rod and the tip broke off.Another reaming rod was opened, and surgery was successfully completed.This (b)(4) will capture the post op event where there was a device removal due to infection.While (b)(4) was created to capture the intra op event in which during the surgery the tip of a 2.5 reaming rod and the tip broke off.This is report 1 of 7 for (b)(4).This report is for an unknown tibial nail.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H4, h6: manufacturing location: monument, manufacturing date: august 20, 2021, expiration date: july 31, 2031, part: 04.034.433s, 10mm ti cann tibial nail - ex w/prox bend 315mm ¿ sterile, lot: 331p744 (sterile).Production order traveler met all inspection acceptance criteria.Inspection sheet, in process / inspect dimensional / final met all inspection acceptance criteria.Packaging label log (pll) was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Sterilization control number (scn) supplied by ethicon (abq) was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removal due to possible infection¿ does not indicate breakage of the nail.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM TI CANN TIBIAL NAIL-EX W/PROX BEND 315MM-STERILE
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14456876
Report Number2939274-2022-01804
Device Sequence Number4685395
Product Code JDS
UDI-Device Identifier10886982095220
UDI-Public(01)10886982095220
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K040762
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.034.443S
Device Catalogue Number04.034.443S
Device Lot Number331P744
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer06/17/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date06/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: 5.0 MM LOCKING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
Patient SexUnknown
Patient Weight86 KG
-
-