• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® SODIUM HEPARINN (NH) BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® SODIUM HEPARINN (NH) BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367874
Medical Device Problem Code Coagulation in Device or Device Ingredient (1096)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® sodium heparinn (nh) blood collection tubes, the device experienced clotting.The following information was provided by the initial reporter.The customer stated: it was reported that tubes are clotting by the time they are being processed by the lab.
 
Additional Manufacturer Narrative
H6: investigation summary bd had not received samples or photos for evaluation.Therefore, retention samples of the incident lot were selected from bd inventory for evaluation and upon completion, no issues relating to clotting were observed.No difficulties were encountered during blood collection and all tubes appeared to exhibit proper fill.Bd was unable to confirm the customer¿s indicated failure (fibrin strand(s)/fibrin mass/clotting) because the defect was not evident in the testing of the complaint lot samples.All samples (retain and control lots) demonstrated clinically acceptable performance for all visual observations evaluated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode clotting.Bd was not able to identify a root cause for the indicated failure mode.H3 other text : see h10.
 
Event or Problem Description
It was reported when using the bd vacutainer® sodium heparinn (nh) blood collection tubes, the device experienced clotting.The following information was provided by the initial reporter.The customer stated: it was reported that tubes are clotting by the time they are being processed by the lab.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SODIUM HEPARINN (NH) BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14458463
Report Number1917413-2022-00284
Device Sequence Number4685984
Product Code JKA
UDI-Device Identifier00382903678747
UDI-Public00382903678747
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K944566
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2023
Device Catalogue Number367874
Device Lot Number1228119
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-