During inspection and testing, the distal end of the scope was loose due to handling.The reported event (markings faded/peeling) was confirmed.The coating of the insertion part was peeling due to handling.In addition, the scope leaked due to holes in the bending section and loose parts on the distal end caused by handling, there were image abnormalities due to peeling of the adhesive on the objective lens, the universal cord and video cable were crushed, and there were scratches on multiple parts of the device.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to stress, handling, or other factors.Olympus will continue to monitor field performance for this device.
|