• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. RHINO-LARYNGO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. RHINO-LARYNGO VIDEOSCOPE Back to Search Results
Model Number ENF-V3
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
During inspection and testing, the distal end of the scope was loose due to handling.The reported event (markings faded/peeling) was confirmed.The coating of the insertion part was peeling due to handling.In addition, the scope leaked due to holes in the bending section and loose parts on the distal end caused by handling, there were image abnormalities due to peeling of the adhesive on the objective lens, the universal cord and video cable were crushed, and there were scratches on multiple parts of the device.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
The customer reported the markings were fading and peeling from the scope.The subject device was sent to an olympus service center for evaluation.During inspection and testing, the distal end of the scope was loose.This report is being submitted for the malfunction found during evaluation (loose parts).There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to stress, handling, or other factors.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHINO-LARYNGO VIDEOSCOPE
Type of Device
RHINO-LARYNGO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14458558
MDR Text Key300334304
Report Number8010047-2022-08555
Device Sequence Number1
Product Code EOB
UDI-Device Identifier04953170411366
UDI-Public04953170411366
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K061313
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberENF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/21/2022
Initial Date FDA Received05/20/2022
Supplement Dates Manufacturer Received06/07/2022
Supplement Dates FDA Received07/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-