MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION CNS-9700A; CENTRAL MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION CNS-9700A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-9700A
Medical Device Problem Codes Device Alarm System (1012); No Audible Alarm (1019); Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
The nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.
 
Additional Manufacturer Narrative
The nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Additional Manufacturer Narrative
Investigation summary: the nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.Investigation conclusion: customer reported that their cns was not alarming for nibp.Nurse indicated that they had an option to turn off the alarm.This suggests that there was the cns alarmed visually.The issue is likely regarding no sound or audible alarm.In a follow-up, customer indicated that the issue has been resolved.But the customer did not know how the issue was resolved.A review of the history of serial number identified no similar events.Based on the available information, a definitive root cause could not be identified.Audio alarms may not be audible when the volume is muted or is set to low volume, the audio cable is unplugged, or when the wrong audio output device is selected.If the audio settings of the device are correct and the cable is properly plugged, audio issues may still occur due to physical damage on the cable.Cable damage may occur due to normal wear and tear or due to human factors.Cables that are often disconnected and reconnected or are positioned in areas where they are in possible with other objects or persons are more susceptible to physical damage.Possible root causes for cable damage are normal wear and tear or mishandling.Correction: b2 required intervention to prevent permanent impairment/damage was marked in error in the initial mdr submitted and the check mark has been removed from the box.
 
Event or Problem Description
The nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-9700A
Common Device Name
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa,, saitama 359-8-580
JA   359-8580
MDR Report Key14458706
Report Number8030229-2022-02876
Device Sequence Number17324012
Product Code MHX
UDI-Device Identifier04931921000663
UDI-Public04931921000663
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K023475
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNS-9700A
Device Catalogue NumberMU-971RA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/11/2006
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-