| Model Number |
CNS-9700A |
| Medical Device Problem Codes |
Device Alarm System (1012); No Audible Alarm (1019); Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.Investigation conclusion: customer reported that their cns was not alarming for nibp.Nurse indicated that they had an option to turn off the alarm.This suggests that there was the cns alarmed visually.The issue is likely regarding no sound or audible alarm.In a follow-up, customer indicated that the issue has been resolved.But the customer did not know how the issue was resolved.A review of the history of serial number identified no similar events.Based on the available information, a definitive root cause could not be identified.Audio alarms may not be audible when the volume is muted or is set to low volume, the audio cable is unplugged, or when the wrong audio output device is selected.If the audio settings of the device are correct and the cable is properly plugged, audio issues may still occur due to physical damage on the cable.Cable damage may occur due to normal wear and tear or due to human factors.Cables that are often disconnected and reconnected or are positioned in areas where they are in possible with other objects or persons are more susceptible to physical damage.Possible root causes for cable damage are normal wear and tear or mishandling.Correction: b2 required intervention to prevent permanent impairment/damage was marked in error in the initial mdr submitted and the check mark has been removed from the box.
|
| |
|
Event or Problem Description
|
|
The nurse reported that they are not getting alarms from the nibp on the central nurse's station (cns).They stated that they are not getting any nibp alarms for patients above or below the set numbers of 170 / 80.They are requesting to have a clinical review of the alarm setting on this cns.One of the nurses in icu informed her that they do have the option to turn the alarms off.The nurse also stated that they would call up to the other departments to verify if they are changing these setting and reach back out to us.Tech support (ts) called for additional complaint information and was told all they know was that the issue is resolved, and they do not know how it was resolved.No patient harm was reported.
|
| |
|
Search Alerts/Recalls
|