| Model Number |
500-55140 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was brought back to the catheterization lab for catheter replacement.A 106x40cm ekosonic catheter was selected for use in a 4 hour left arterial unilateral case.Tissue plasminogen activator (tpa) was given at 25ml/hr and coolant was at 35ml/hr.Difficulty placing the catheter in the left leg occurred.An all groups disabled alarm displayed.The nurse was in the intensive care unit (icu) with the patient and could not get the control unit (cu) to run.Troubleshooting was performed; however, the alarm persisted.The patient was brought back to the catheterization lab for catheter replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device evaluated by mfr: the returned device was inspected for any damage or irregularities.The infusion catheter (ic) was returned with blood in the coolant lines to the luer.The ultrasonic core (usc) device showed damage in the form of a severe kink at the luer barb and at the distal tip.The usc had advancement tool markings/ kinks throughout the device.The ic showed some damage in the form of wavy wires throughout the device.The device ran in a warm water bath to attempt to replicate the sites reported alarm.About 15mins into treatment the device began alarming with an e311 alarm.Disconnecting and reconnecting the device resolved the issue but the device alarmed about 30sec later with the same alarm.The device continued this type of alarming until the device was shut down.Analysis of the in-house event log showed that the deice usc phase was too high, so the groups shut off.This is likely due to wire damage of the usc.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was brought back to the catheterization lab for catheter replacement.A 106x40cm ekosonic catheter was selected for use in a 4 hour left arterial unilateral case.Tissue plasminogen activator (tpa) was given at 25ml/hr and coolant was at 35ml/hr.Difficulty placing the catheter in the left leg occurred.An all groups disabled alarm displayed.The nurse was in the intensive care unit (icu) with the patient and could not get the control unit (cu) to run.Troubleshooting was performed; however, the alarm persisted.The patient was brought back to the catheterization lab for catheter replacement.
|
| |
|