Model Number 9-ASD-008
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2022, an 8mm amplatzer septal occluder (7443783) was chosen for implant using a 9f amplatzer trevisio intravascular delivery system (8287845).During deployment of the device in the left atrium, the distal disc appeared as a cobra deformation.The device was recaptured and removed from the patient.The device was able be reformed to its proper shape outside the body.The device was deployed in the patient a second time, and the device took a cobra deformation again.There was no interaction with cardiac structures during deployment, and no angulation was noticed in the delivery system.The device was exchanged with a 10mm amplatzer septal occluder (8031289), and the delivery system was exchanged with a 10f amplatzer trevisio intravascular delivery system (8115344).The replacement device was successfully implanted.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient has since been discharged.
 
Additional Manufacturer Narrative
An event of device deformity was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key14459689
Report Number2135147-2022-00301
Device Sequence Number4482587
Product Code MLV
UDI-Device Identifier05415067019165
UDI-Public05415067019165
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model Number9-ASD-008
Device Catalogue Number9-ASD-008
Device Lot Number7443783
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMPLATZER TREVISIO INTRAVASCULAR DS, 9-ATV09F45/80
Patient SexMale
-
-