• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (US); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (US); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW 7 XB BISECTOR (US)
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Trackwise id (b)(4).
 
Event or Problem Description
Related to (b)(4).The hospital reported that during an endoscopic vein harvesting procedure,vasoview 7 xb malfunctioned (cautery not working).Vein was good and no excess bleeding.A replacement device was used to complete the procedure.No harm was done to the patient.
 
Additional Manufacturer Narrative
Trackwise id #(b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).The device was returned to the factory for evaluation on 05/19/2022.An investigation was conducted on 07/05/2022.A visual inspection was conducted.Signs of clinical use and evidence of blood and charred material was observed.The electrodes were observed to be intact, with no visual defects observed.A pre-cautery test was performed and repeated 10 times over a 10 minute period according to the procedure in the instructions for use (cv000001599).The device failed the pre-cautery test with a reference generator (recommended setting at 18) and reference bipolar cord.An activation test was performed and the device did not activate.The bipolar cord connection was manipulated during activation.The device did not activate.The device did not produced steam and the saline did not ¿boil¿ on the test gauze each time.Based on the returned condition of the device, the reported failure "failure to deliver energy" was confirmed.The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event or Problem Description
Related to (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW 7 XB BISECTOR (US)
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key14459714
Report Number2242352-2022-00430
Device Sequence Number14611774
Product Code GEI
UDI-Device Identifier00607567700369
UDI-Public00607567700369
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K041981
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2023
Device Model NumberVASOVIEW 7 XB BISECTOR (US)
Device Catalogue NumberC-VH-3200
Device Lot Number25161194
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer05/19/2022
07/13/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date06/15/2022
07/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
-
-