• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. MARQUEE; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. MARQUEE; BIOPSY INSTRUMENT Back to Search Results
Model Number MQ1210
Medical Device Problem Codes Break (1069); Self-Activation or Keying (1557); Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturing location was selected as unknown due to system limitations.The manufacturing location for the mission product is bd (b)(4).As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 11/2024).
 
Event or Problem Description
It was reported that during a biopsy procedure, the device allegedly exploded when it was being primed.It was further reported that the tip was allegedly do not feel as sharp.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: the manufacturing location was selected as unknown due to system limitations.The manufacturing location for the mission product is bd-santo domingo.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one marquee disposable core biopsy instrument was returned for evaluation.Upon visual evaluation, the device appeared clean throughout, and device was received in its initial position.The penetration depth marker was set to 25mm.No break was noted to the device.And the tip of the needle was noted to be slightly dull.No other visual anomalies were noted to the device.Further functional testing was not performed due to the condition of returned device.Therefore, the investigation is confirmed for the reported blunt needle as the tip of the needle was noted to be slightly dull.However, the investigation is inconclusive for the reported self-activation as the functional testing was not performed due to the condition of returned device.A definitive root cause for the reported blunt needle and self-activation could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 11/2024), g3, h6 (method) h11: h6 (device, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during a biopsy procedure, the device allegedly exploded when it was being primed.It was further reported that the tip was allegedly do not feel as sharp.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARQUEE
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14460652
Report Number2020394-2022-00391
Device Sequence Number4515865
Product Code KNW
UDI-Device Identifier00801741097119
UDI-Public(01)00801741097119
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K133948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMQ1210
Device Catalogue NumberMQ1210
Device Lot Number0001446578
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-