• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TEPHA, INC. - 3005670760 GALAFLEX; BREAST - ABSORBABLE MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TEPHA, INC. - 3005670760 GALAFLEX; BREAST - ABSORBABLE MESH Back to Search Results
Model Number 100035-01
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 06/06/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Per clinical trial (b)(6): on (b)(6) 2017, patient underwent bilateral cosmetic augmentation mastopexy using a full vertical and horizontal t incision.1 piece of galaflex was used.There were no concomitant procedures done.On (b)(6) 2018, an ae of 'reoperation' was reported.On (b)(6) 2019, patient had bilateral capsulotomy with breast implant replacement.Resolved without sequelae on (b)(6) 2019.Review by medical affairs identifies reoperation as likely related to the device and serious.
 
Additional Manufacturer Narrative
Based on the information provided, no conclusions can be made.As reported, patient experienced the post-op complication of capsular contracture and had additional surgery.Medical affairs has assessed the patient¿s postoperative course as being serious and likely related to the study device.Review of the device history records confirmed the device was manufactured to specification.Note: this mdr report is related to a retrospective study conducted to proactively gather real-world evidence on the use and safety of the p4hb product family, including galaflex scaffold, galaflex 3d, and galaflex 3dr.All safety events from the study are adjudicated by the bd medical monitor for an independent assessment of severity and relatedness.In an abundance of caution, events for which relatedness cannot be ruled out are processed for mdr reporting.Not returned - remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAFLEX
Common Device Name
BREAST - ABSORBABLE MESH
Manufacturer (Section D)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer (Section G)
TEPHA, INC. - 3005670760
99 hayden ave suite 360
lexington MA 02421
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14460903
Report Number1213643-2022-00258
Device Sequence Number4506893
Product Code OOD
UDI-Device Identifier00855920006010
UDI-Public(01)00855920006010
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K140533
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Model Number100035-01
Device Catalogue NumberGP0206
Device Lot Number160186
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2022
Initial Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/22/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age23 YR
Patient SexFemale
Patient Weight52 KG
-
-