• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. LINER 10 DEGREE ELEVATED RIM 3.5 MM OFFSET 36 MM I.D.; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. LINER 10 DEGREE ELEVATED RIM 3.5 MM OFFSET 36 MM I.D.; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Osteolysis (2377); Osteopenia/ Osteoporosis (2651)
Date of Event 04/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a hip revision approximately 20 to 30 years later due to poly wear and lysis.Subsequently, patient was revised approximately three (3) years later due to fracture, lysis, and pelvic discontinuity.Unfortunately the bone loss was too much for the tmars construct to support.A custom cage was created to reconstruct the acetabulum.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: cat#00-4894-058-10 acet agmt sz 58 x 10 lot#63129477, cat#00-6250-065-25 bone scr 6.5x25 self-tap lot#64206965, cat#00-6250-065-40 bone scr 6.5x40 self-tap lot#64005644, cat#00-6250-065-50 bone scr 6.5x50 self-tap lot#63451417, and cat#00-6624-065-50 bone scr 6.5x50 self-tap lot#63759097.Foreign country: (b)(6).The device will not be returned for analysis, per hospital policy; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2022 - 01503.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4: udi number; h2; h3; h6.Visual examination of the provided picture identified the shell explanted and covered in bio-debris.No further evaluation can be made from the provided picture.Pictures were not provided of the liner.Medical records were not provided.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.It was identified that zimmer biomet devices were implanted with competitor devices.Zimmer biomet has not confirmed the compatibility for these combinations of devices.It is unknown if these off-label usages may have caused or contributed to the reported events.A definitive root cause cannot be identified.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER 10 DEGREE ELEVATED RIM 3.5 MM OFFSET 36 MM I.D.
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14460904
Report Number0001822565-2022-01502
Device Sequence Number4506894
Product Code LPH
UDI-Device Identifier00889024124868
UDI-Public(01)00889024124868(17)230731(10)64095426
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K972774
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberN/A
Device Catalogue Number00631006036
Device Lot Number64095426
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight80 KG
-
-