• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT Back to Search Results
Model Number AVSM10100
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 02/2023).
 
Event or Problem Description
It was reported that during a stent graft placement procedure in the upper arm fistula, allegedly resistance was felt at the beginning and after deploying a certain length, the deployment wheel turned without further covered stent deployment.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not returned for evaluation and photos were not provided for review.Based on information available, the investigation is inconclusive for partial deployment.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.With regards to preparation and accessories, the instructions for use states "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated", "flush the delivery system through the luer port at the proximal end of the handle with sterile saline until the saline exits the tip of the system" and "0.035 inch guidewire of appropriate length, introducer sheath with appropriate inner diameter".Regarding correct deployment techniques, the instructions for use further states that "with your free hand, maintain a stationary hold on the white stability sheath during covered stent deployment and adjust for placement accuracy if necessary.Hold the white stability sheath as close as possible to the introducer without touching the dark brown moving catheter of the distal catheter assembly.Maintain the remainder of the white stability sheath".H10: d4 (expiry date: 02/2023), g3.H11: h6 (result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a stent graft placement procedure in the upper arm fistula via bareback, the device allegedly partially deployed.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA VASCULAR COVERED STENT
Common Device Name
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14460964
Report Number9681442-2022-00156
Device Sequence Number14627631
Product Code PFV
UDI-Device Identifier00801741106415
UDI-Public(01)00801741106415
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P170042
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAVSM10100
Device Catalogue NumberAVSM10100
Device Lot NumberANFP0822
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer07/06/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient RaceWhite
-
-