• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAX MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955626
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Tachycardia (2095)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that during continuous renal replacement therapy on a prismax machine, the operator erroneously programmed the patient fluid removal rate (pfr) as the flow rate prescribed for the dialysate flow rate (dfr).The patient fluid removal rate was inadvertently set to 1500 ml/hr instead of 50 ml/hr.The error was identified after four hours into treatment and the patient became hypotensive and tachycardic.The effective ultrafiltration volume removed from the patient was approximately 6l over a four hour period (1500 ml/hr x 4).The error was corrected and the patient was treated with multiple vasoactive medications, sodium and calcium (intravenously), saline (1000 ml), lactated ringer (250 ml) and albumin.No additional information is available.
 
Additional Manufacturer Narrative
The device was not received for evaluation; therefore, a device analysis could not be completed.Based on the reported information, the reported condition was verified that the fluid removal was due to an erroneous setting by the operator.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAX MACHINES
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - BROOKLYN PARK
7601 northland drive
n suite 170
minneapolis MN 55428
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14461351
Report Number3003504604-2022-00014
Device Sequence Number14633298
Product Code KDI
UDI-Device Identifier00085412639499
UDI-Public(01)00085412639499
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K190910
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number955626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexUnknown
-
-