Catalog Number 8065751614
Device Problem Fluid/Blood Leak (1250)
Patient Problem Retinal Tear (2050)
Event Date 05/12/2022
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that one of the trocars had a leakage and part of the retina was trapped in the trocar resulting in the creation of another hole on the retina apart from the existing retinal hole due to trauma during the vitrectomy procedure.The surgeon successfully managed to release the retina from the trocar.The surgeon decided to continue the procedure with the same cassette due to the fact that the patient had a trauma.The procedure was completed successfully.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.Investigations have been completed and manufacturing process enhancements have been implemented in order to improve the performance of the valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14461479
MDR Text Key292369348
Report Number1644019-2022-00388
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657516148
UDI-Public00380657516148
Combination Product (y/n)N
Reporter Country CodeCY
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number8065751614
Device Lot Number2531981H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM; CONSTELLATION VISION SYSTEM
Patient Outcome(s) Other;
Patient Age57 YR