| Model Number |
1320-0200 |
| Medical Device Problem Codes |
Device Difficult to Setup or Prepare (1487); Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Implant Pain (4561)
|
| Date of Event |
04/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the gamma lag screw was being revised, due to patient discomfort, and the surgeon was unable to thread the inner connecting piece onto the lag screw itself.A peel pack was available was used instead.Everything was able to attach afterwards.Upon inspection of the instrument in question, it was noticed that the threads were damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note correction to d9/h3 as the device was not returned for evaluation and correction of h6 clinical code the reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the gamma lag screw was being revised, due to patient discomfort, and the surgeon was unable to thread the inner connecting piece onto the lag screw itself.A peel pack was available was used instead.Everything was able to attach afterwards.Upon inspection of the instrument in question, it was noticed that the threads were damaged.
|
| |
|