• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT; DIALYSIS CATHETER Back to Search Results
Model Number 5733694
Medical Device Problem Codes Restricted Flow rate (1248); Suction Problem (2170)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/03/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 09/2021).
 
Event or Problem Description
It was reported through results of a clinical trial that sometime post index procedure completion, device failure noted.Post insertion attempted dialysis, unable to use due to no pulls or sluggish flow from venous port and was not successfully treated with hemosplit catheter due to dysfunction.Approximately seven months and ten days post index procedure completion, the patient was expired.
 
Event or Problem Description
It was reported through results of a clinical trial that sometime post index procedure completion, device failure noted.Post insertion attempted dialysis, unable to use due to no pulls or sluggish flow from venous port and was not successfully treated with hemosplit catheter due to dysfunction.Approximately seven months and ten days post index procedure completion, the patient was expired.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation cannot be confirmed for the reported aspiration difficulty and low flow issues, as no objective evidence was provided for review.A definitive root cause for the reported flow issues, or the reported patient death, could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 09/2021), g3, h6 (device, method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPLIT
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14461860
Report Number3006260740-2022-01871
Device Sequence Number20004956
Product Code MSD
UDI-Device Identifier00801741013119
UDI-Public(01)00801741013119
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K030020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5733694
Device Catalogue Number5733694
Device Lot NumberREDY2508
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexFemale
Patient Weight70 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-