AVANOS MEDICAL INC. ON-Q* C-BLOC WITH SELECT-A-FLOW AND ONDEMAND, 400 ML X 1-7 +5 ML BOLUS/60 MIN; ELASTOMERIC - COMBO
|
Back to Search Results |
|
| Model Number |
CB005 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Fill volume: 550 ml, flow rate: 6ml/hr, procedure: rotator cuff, cathplace: unknown, date of procedure: 18-apr-2022.It was reported the pump "infused entire amount (550ml) in a day and a half." there was no reported injury.Additional information received 29-apr-2022 stated the "patient did not suffer any adverse side effects.No signs or symptoms of local anesthetic toxicity.Patient is currently stable and at home recovering from surgery per expected outcomes.".
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for the reported lot number, 30156196, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.The product involved in the report has been returned and is being processed for evaluation.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
All information reasonably known as of 01-jun-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was evaluated for flow and pressure testing.All tested rates met specification with a +/-20% tolerance.No device problem was found.Root cause could not be determined.All information reasonably known as of (b)(6) 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint comp (b)(6).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|
|
|