| Catalog Number |
2H8603 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified quantity of a clearlink system non-dehp extension sets were leaking from the side of the port.The issue was identified during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction b5: as a result of the event, the infusion was stopped and a new one was placed and restarted (omitted on initial).H10: the actual devices were not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.Functional testing was performed including clear passage and pressure testing; and the results were satisfactory.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|