Model Number 6937A-35
Medical Device Problem Codes Fracture (1260); High impedance (1291); Device Markings/Labelling Problem (2911); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: product id: cd2311-36 competitor icd, date of implant: (b)(6) 2013; if information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead triggered an alert for high superior vena cava (svc) coil impedance, that the lead had also had variances in its svc coil impedance and that the lead was fractured.The lead was reprogrammed and remains in use.It was additionally noted that the lead is being used off-label.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSVENE CS/SVC LEAD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14462781
Report Number2649622-2022-09815
Device Sequence Number17314990
Product Code LWS
UDI-Device Identifier00613994492388
UDI-Public00613994492388
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P980050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/27/2014
Device Model Number6937A-35
Device Catalogue Number6937A-35
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Initial Report FDA Received Date05/20/2022
Date Device Manufactured10/10/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6937A-35 LEAD, 4968-25 LEAD, 4968-25 LEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient SexMale
-
-