| Model Number |
71173503 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during surgery, two (2) 2.7mm drill {} bit w/qc broke and one 2.7mm drill {} bit w/qc lost its edge, when drilling into the cement to place the screws.Surgery was finished, without any delay, with a sn back-up device.No harm to the patient or further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the devices were not returned for evaluation and the reported event could not be confirmed.A review of complaint history of the previous 12 months revealed similar events for the listed devices, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.These are single use devices, damage from misuse, excessive cleaning or rough handling are likely probable causes of the reported event.The contribution of the devices to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual products involved, our investigation could not proceed.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that, during surgery, two (2) 2.7mm drill {} bit w/qc broke, when drilling into the cement to place the screws.Surgery was finished, without any delay, with a sn back-up device.No harm to the patient or further complications reported.
|
| |
|