|
According to the reporter, on an open low anterior resection, while in the colon, the surgeon performed full, complete squeezes of the handle but no staples were deployed.The surgeon did not cut the tissue.Another loading unit was used to complete the case.There was no patient injury.
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the staples did not deploy from the stapler after the handle was squeezed.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: squeeze the handle to begin approximating the tissue.There is a tactile ¿pre-clamp¿ position built into the instrument where the handle can be released without returning to its original position to allow for final positioning of the tissue within the jaws.After the tissue is properly positioned continue the handle stroke until the instrument is clamped.The handle will return to the original position.To fire the instrument squeeze then handle a second time until it reaches a solid stop and it locks in the back position.When fully fired, the handle will remain in the locked position until the black release button is pushed, and the instrument opened.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|