U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6530
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 85-95% stenosed target lesion was located in the left anterior descending artery.A 8mm x 3.50mm nc quantum apex balloon catheter was advanced for dilatation.However, during the procedure, the shaft was reported to be damaged.The procedure was completed with a different device.There were no patient complications reported.It was further reported that the shaft was broken.
 
Event or Problem Description
It was reported that shaft break occurred.The 85-95% stenosed target lesion was located in the left anterior descending artery.A 8mm x 3.50mm nc quantum apex balloon catheter was advanced for dilatation.However, during the procedure, the shaft was reported to be damaged.The procedure was completed with a different device.There were no patient complications reported.It was further reported that the shaft was broken.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of an nc quantum apex balloon catheter.The device was visually and microscopically examined.At 45cm from the strain relief, the hypotube of the device was kinked.There was a separation of the hypotube 46.6cm from the strain relief.The separated ends were jagged and ovaled in shape indicating the device was kinked prior to separation.At 1.3cm distal from the separation, the hypotube was kinked.The balloon was tightly folded and there was no evidence of fluid within the device.The tip of the device was severely damaged.Product analysis confirmed the reported break as there was separation to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14463230
Report Number2134265-2022-05766
Device Sequence Number4904227
Product Code LOX
UDI-Device Identifier08714729783176
UDI-Public08714729783176
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/25/2025
Device Model Number6530
Device Catalogue Number6530
Device Lot Number0028748415
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/10/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date05/20/2022
Supplement Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-