| Model Number |
04V5121 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The literature article by melville, anna and thomas, sunethra devika c ¿paraprotein interference of automated total bilirubin measurement¿ documented paraprotein interference in the alinity c total bilirubin assay.The author documented falsely depressed alinity c total bilirubin results generated on the alinity c instrument for a (b)(6) male patient.The initial result was <9 umol/l (reference interval 4 to 20 umol/l), all other routine results were normal except total protein 120 g/l (reference interval 60 to 85 g/l), albumin 38 g/l (reference interval 39 to 50 g/l) and globulin 82 g/l (reference interval 23 to 39 g/l).The patient¿s previous results for bilirubin were <9 umol/l on several occasions, however, 12 months prior the bilirubin result on roche platform was 54 umol/l, other results total protein 116 g/l (reference interval 66 to 83 g/l), albumin 37 g/l and globulin 79 g/l).Additionally, the findings of the study documented that the presence of paraprotein can affect the measured concentration of bilirubin using the abbott alinity assay.Furthermore, igm appears to exert a more significant effect on bilirubin measurement.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
Data and information provided was reviewed and support the complaint issue.Return testing was not performed as returns were not available.Device history record review was performed on list 04v51, which did not show any potential non-conformances, deviations, or non-conformances.Trending review did not identify any trends for the complaint issue.Labeling was reviewed and found to adequately address the issue under review.Labeling in the total bilirubin package insert and the alinity ci series operations manual advises the operator and physician to review all results from the patient along with clinical data, other tests, symptoms and clinical evaluation to properly evaluate the applicability of test results.The findings of the study documented that the presence of paraprotein can affect the measured concentration of bilirubin using the abbott alinity assay.Based on the information within the complaint record, the device met performance specifications and performed as intended at the customer site.A deficiency was not identified.
|
| |
|
Search Alerts/Recalls
|
|
|