CORDIS CORPORATION SABER 3.5MM15CM 150; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
| Model Number |
48003515X |
| Medical Device Problem Code |
Partial Blockage (1065)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|
Event or Problem Description
|
|
As reported, the saber 3.5mm 15cm 150 had something stuck inside the balloon; therefore, the non-cordis wire could not pass through the device.The case was completed with a new cordis balloon catheter.There was no reported injury to the patient.The product was stored properly according to the instructions for use (ifu).There was no difficulty removing the product from the hoop.There was no difficulty removing the protective balloon cover, stylet, or any of the sterile packaging components.The device was never inserted into the patient.The device was attempted to be prepped per the ifu.The obstruction was noticed while pre-loading outside the patient onto the non-cordis guidewire.The catheter was never in an acute bend and the catheter was never kinked.The device will not be returned for evaluation as multiple attempts were made without success.
|
| |
|
Event or Problem Description
|
|
As reported, the saber 3.5mm 15cm 150 had something stuck inside the balloon; therefore, the non-cordis wire could not pass through the device.The case was completed with a new cordis balloon catheter.There was no reported injury to the patient.The product was stored properly according to the instructions for use (ifu).There was no difficulty removing the product from the hoop.There was no difficulty removing the protective balloon cover, stylet, or any of the sterile packaging components.The device was never inserted into the patient.The device was attempted to be prepped per the ifu.The obstruction was noticed while pre-loading outside the patient onto the non-cordis guidewire.The catheter was never in an acute bend and the catheter was never kinked.The device will not be returned for evaluation as multiple attempts were made without success.
|
| |
|
Additional Manufacturer Narrative
|
|
As reported, the saber 3.5mm 15cm 150 had something stuck inside the balloon; therefore, the non-cordis wire could not pass through the device.The case was completed with a new cordis balloon catheter.There was no reported injury to the patient.The product was stored properly according to the instructions for use (ifu).There was no difficulty removing the product from the hoop.There was no difficulty removing the protective balloon cover, stylet, or any of the sterile packaging components.The device was never inserted into the patient.The device was attempted to be prepped per the ifu.The obstruction was noticed while pre-loading outside the patient onto the non-cordis guidewire.The catheter was never in an acute bend and the catheter was never kinked.The device was not returned for evaluation as multiple attempts were made without success.A product history record (phr) review of lot 82183812 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿guidewire lumen- obstructed - during prep¿ was not confirmed as the device was not returned for analysis.The exact cause could not be determined.It is likely procedural factors and handling of the device may have contributed to the event reported by the customer.Use of the device with the concomitant hydrophilic wire may also be a contributing factor.However, without the return of the device for analysis or concomitant wire used with the device, it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.Flush all devices entering the vascular system with sterile heparinized saline or similar isotonic solution.Prior to use, ensure all devices have been flushed and air is removed from the system according to standard medical practice.Failure to do so could result in air entering the vascular system.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
|
| |
|
Search Alerts/Recalls
|
|
|