• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT W/BIOPATCH PC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT W/BIOPATCH PC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a bent stylet is confirmed; however, the exact cause is unknown.One 3cg stylet was returned for evaluation.An initial visual observation showed some use residue on the returned sample.Bends were observed in the returned stylet approximately 4.4 cm proximal to its distal tip as well as approximately 7 cm and 14 cm distal to the yellow handle of the stylet.No breaks were observed in the core wire of the stylet and the distal tip was found to be present and intact.A microscopic observation revealed no breaks or tears in the tubing at the distal end of the stylet, and the bend within this tubing was found to be near the interfacing of two magnets.Insufficient evidence was found on the returned sample to indicate a specific root cause of the bend in the stylet; however, possible contributing factors include advancement of the stylet against resistance and damage during manipulation, handling, or use.A lot history review (lhr) of refv2196 showed two other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported via medwatch, "they have been checking stylets over when removing the piccs.Noticed a bend in wire but no difficulties with insertion." additional info.Received 12/15/2021: what type of catheter securement was used? statlock.On 04/27/2022 - the returned sample exhibited multiple bends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT W/BIOPATCH PC
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14465316
Report Number3006260740-2022-01910
Device Sequence Number14594882
Product Code LJS
UDI-Device Identifier00801741155123
UDI-Public(01)00801741155123
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1395108D4
Device Lot NumberREFV2196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2022
Initial Report FDA Received Date05/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age46 YR
Patient SexFemale
-
-