SYNTHES GMBH POSITIONER F/AIMING BLOCK; ORTHOPAEDIC IMPLANT AIMING/GUIDING BLOCK, REUSABLE
|
Back to Search Results |
|
| Catalog Number |
03.108.006 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
|
| |
|
Event or Problem Description
|
|
Device report from depuy synthes reports an event in colombia as follows: this is report 1 of 2 for (b)(4).It was reported by the affiliate in colombia that during an unknown orthopedic procedure on (b)(6) 2022, it was observed that one of the holes in the guide did not allow the passage of the needle indicated for the procedure while the system locking screw was rolled.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This complaint involves two (2) devices.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: d10 h3, h6: a photo investigation was completed: visual analysis of the provided photo revealed that there was no damage or defects with the aiming block.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as device was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the cap screw of the positioner f/aiming block has loose, the screw not have a tight fit causing incorrect functionality with the positioning plate.The assemble issue was not confirmed due to the positioner f/aiming block and the mating device were assemble correctly.No other problems were reported.A dimensional inspection was not performed for the positioner f/aiming block due to the design of the device.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the positioner f/aiming block would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: part # 03.108.006, lot # l657358: a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.
|
| |
|