MAUDE Adverse Event Report: CORDIS CORPORATION SABER 6MM30CM 150; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SABER 6MM30CM 150; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 48006030X
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the three saber 6mm 30cm balloons busted in the superficial femoral artery (sfa) when being inflated to recommended inflation parameters.All three devices were removed intact (in one piece) from the patient.The case was completed with the use of a fourth saber 6mmx30cm 150.There was no reported injury to the patient.The products were stored properly according to the instructions for use (ifu).There was difficulty noted removing the products from the hoop.There was no difficulty removing the protective balloon covers, stylets, or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting any of the devices into the patient.All three devices were prepped per the ifu and were able to maintain negative pressure.The lesion was in the superficial femoral artery.The lesion had moderate calcification with little vessel tortuosity.It was a chronic total occlusion (cto) to be treated.The balloons did not rupture while in a stent.The non-cordis inflation device was filled with 50/50 saline/contrast ratio.The inflation device was used successfully with other devices during the case.There was no resistance/friction while inserting the balloons through the rotating hemostatic valve or while inserting them through the guide catheter.The three balloons advanced through the vessel without difficulty.The balloons had slight difficulty while crossing the lesion.The catheters were never in an acute bend and were never kinked while being used.The devices will be returned for evaluation.
 
Additional Manufacturer Narrative
This device is available for analysis, but the engineering report is not yet available.However, it will be submitted within 30 days upon receipt.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the three saber 6mm 30cm balloons busted in the superficial femoral artery (sfa) when being inflated to recommended inflation parameters.All three devices were removed intact (in one piece) from the patient.The case was completed with the use of a fourth saber 6mmx30cm 150.There was no reported injury to the patient.The products were stored properly according to the instructions for use (ifu).There was difficulty noted removing the products from the hoop.There was no difficulty removing the protective balloon covers, stylets, or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting any of the devices into the patient.All three devices were prepped per the ifu and were able to maintain negative pressure.The lesion was in the superficial femoral artery.The lesion had moderate calcification with little vessel tortuosity.It was a chronic total occlusion (cto) to be treated.The balloons did not rupture while in a stent.The non-cordis inflation device was filled with 50/50 saline/contrast ratio.The inflation device was used successfully with other devices during the case.There was no resistance/friction while inserting the balloons through the rotating hemostatic valve or while inserting them through the guide catheter.The three balloons advanced through the vessel without difficulty.The balloons had slight difficulty while crossing the lesion.The catheters were never in an acute bend and were never kinked while being used.The devices will be returned for evaluation.
 
Event or Problem Description
As reported, the three saber 6mm 30cm balloons busted in the superficial femoral artery (sfa) when being inflated to recommended inflation parameters.All three devices were removed intact (in one piece) from the patient.The case was completed with the use of a fourth saber 6mmx30cm 150.There was no reported injury to the patient.The products were stored properly according to the instructions for use (ifu).There was difficulty noted removing the products from the hoop.There was no difficulty removing the protective balloon covers, stylets, or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting any of the devices into the patient.All three devices were prepped per the ifu and were able to maintain negative pressure.The lesion was in the superficial femoral artery.The lesion had moderate calcification with little vessel tortuosity.It was a chronic total occlusion (cto) to be treated.The balloons did not rupture while in a stent.The non-cordis inflation device was filled with 50/50 saline/contrast ratio.The inflation device was used successfully with other devices during the case.There was no resistance/friction while inserting the balloons through the rotating hemostatic valve or while inserting them through the guide catheter.The three balloons advanced through the vessel without difficulty.The balloons had slight difficulty while crossing the lesion.The catheters were never in an acute bend and were never kinked while being used.The devices will be returned for evaluation.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections have been updated accordingly: b5, g3, h1, h2, h3 and h6.As reported, the three saber 6mm x 30cm balloons busted in the superficial femoral artery (sfa) when being inflated to recommended inflation parameters.All three devices were removed intact (in one piece) from the patient.The case was completed with the use of a fourth saber 6mmx30cm 150.There was no reported injury to the patient.The products were stored properly according to the instructions for use (ifu).There was difficulty noted removing the products from the hoop.There was no difficulty removing the protective balloon covers, stylets, or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting any of the devices into the patient.All three devices were prepped per the ifu and were able to maintain negative pressure.The lesion was in the superficial femoral artery.The lesion had moderate calcification with little vessel tortuosity.It was a chronic total occlusion (cto) to be treated.The balloons did not rupture while in a stent.The non-cordis inflation device was filled with 50/50 saline/contrast ratio.The inflation device was used successfully with other devices during the case.There was no resistance/friction while inserting the balloons through the rotating hemostatic valve or while inserting them through the guide catheter.The three balloons advanced through the vessel without difficulty.The balloons had slight difficulty while crossing the lesion.The catheters were never in an acute bend and were never kinked while being used.A non-sterile saber 6mm x 30cm 150 percutaneous transluminal angioplasty (pta) was received for analysis inside a plastic bag.Per visual analysis, a rupture in the balloon of the unit was observed.No other anomalies found.Per functional analysis, a lab sample inflator/deflator device partially filled with water was attached to the inflation lumen of the unit and pressure was applied.A balloon leakage was observed at a rupture on the balloon surface.No other anomalies were observed.Per microscopic analysis, the unit was inspected under the vision system and the balloon rupture was confirmed on the balloon.Sem analysis revealed the balloon of the saber presented evidence of a ruptured condition and scratch marks near the damaged area.This type of damage is commonly caused during the interaction of the material with a sharp object or mechanical damage.It is very likely that the same factors that caused the observed scratch marks on the balloon probably led to the damaged condition found on the received device.It seems the material near the damage was ruptured with a sharp object from the outside of the device.No other anomalies were observed during the sem analysis.A product history record (phr) review of lot 82239678 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ were confirmed via device analysis as a leakage was noted during functional analysis for all three devices.However, the exact cause cannot be determined.A rupture was noted on the balloon material for all three devices.The outer surfaces of the balloon material presented evidence of scratch marks adjacent to the ruptures noted.It is likely vessel characteristics of moderate calcification and a chronic total occlusion likely contributed to the reported events as evidenced by device analysis.The balloon material near the ruptures, appears to have been torn either due to the interaction of the balloon with calcified spicules located on the lesion or with a sharp object from the outside of the balloon.According to the safety information in the instructions for use, ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER 6MM30CM 150
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key14465551
Report Number9616099-2022-05653
Device Sequence Number14581651
Product Code LIT
UDI-Device Identifier20705032068182
UDI-Public(01)20705032068182(17)241231(10)82239678
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number48006030X
Device Catalogue Number48006030X
Device Lot Number82239678
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer06/24/2022
06/30/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date06/20/2022
06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BOSTON SCIENTIFIC ENCORE-INFLATION DEVICE; BOSTON SCIENTIFIC ENCORE-INFLATION DEVICE
Patient SexUnknown
-
-