MAUDE Adverse Event Report: K2M, INC. AN INSERTER INNER SHAFT; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. AN INSERTER INNER SHAFT; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number 602-90780
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the knob of an aleutian an inserter and inner shaft broke off during intra-operative implant placement.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the inner shaft.
 
Additional Manufacturer Narrative
Visual, dimensional, functional and material analysis could not be performed as the device was discarded by the hospital.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, no adverse trends were identiifed.Limited information regarding the event was made available to stryker.Multiple attempts were made to obtain additional information, but no response was received.A root cause cannot be established with the information available.If additional information is received, the investigation will be reopened and updated.
 
Event or Problem Description
It was reported that the knob of an aleutian an inserter and inner shaft broke off during intra-operative implant placement.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the inner shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AN INSERTER INNER SHAFT
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14465737
Report Number3004774118-2022-00201
Device Sequence Number17302605
Product Code MAX
UDI-Device Identifier10888857157255
UDI-Public10888857157255
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K160547
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number602-90780
Device Catalogue Number602-90780
Device Lot NumberTBC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer07/25/2022
Initial Report FDA Received Date05/21/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-